Quick-Check User Instructions 23
7 Troubleshooting
Problem Possible Cause Solution
Powers on momentarily and
turns of
f
Low battery Replace with high quality alkaline batteries
Does not power on Low battery Replace with high quality alkaline batteries
Batteries installed
backwards or no
spring contact
Insure that positive terminals of both batteries
(nub) face inward – towards the black cap. Check
that spring is attached to the battery cap.
Software reset Remove battery cap & reinstall after one minute.
A
ttempt to turn power on again.
Display contrast too
light
Hold the Right Arrow key down while pressing
the F
2 key several times to increase the display
contrast. If nothing occurs, release both keys.
Press the power button and try again.
Display is completely dark Display contrast too
dark
Hold the Arrow key down while pressing the F1
key several times to d
ecrease the display
contrast.
Display drifts downward
once installed
Wire material is
cre
eping and internal
friction between wires
is relieved.
This is normal behavior of wire. Lower display
resolution to mask this ef
fect.
Temperature not accurate Instrument changed
temperature
environments
Allow instrument to remain in environment until
temp stabilizes or enter temp manually
Instrument exposed
to sun
Enter temp manually